Fluke Calibration
Fluke 1586A Super-DAQ Precision Temperature Scanner
Model No :
Fluke 1586A
Part Code :
4409939

Overview

The 1586A Super-DAQ collects time-stamped precision temperature and electrical measurements for data analysis by technicians, engineers, and quality control personnel to verify process control, analyze interactive systems, ensure conformance to quality standards, or to correlate related events for R&D or troubleshooting. Measurement data and statistics can be viewed in tabular format for all active channels. With the graphing feature, up to four channels can be plotted at the same time, making it easy to quickly assess test setup and results before analyzing the data on a PC.

Features
- Best temperature measurement accuracy
- Flexible configuration for the factory or benchtop
- Multiple modes of operation
- Real-time graphing in color
- Data portability and security
- Automated sensor calibration

Major Applications
- Thermal mapping
- Temperature validation
- Process sensor calibration

Technical Specifications
- Temperature range : -200°C to 1200°C (depending on the sensor)
- Resistance range : 0 O to 4 kO
- Offset compensation : 0 O to 400 O, 4-wire : Automatic current reversal
400 O to 4000 O or 3-wire none
- Source current reversal interval (0 O to 400 O range) : Fast sample rate : 2 ms
- Medium sample rate : 250 ms
- Slow sample rate : 250 ms
- Maximum Lead Resistance (4-wire O) : 2.5% of range per lead for 400 O and 4 kO range
- Thermistor:
• Temperature Range : -200°C to 400°C (depending on the sensor)
• Resistance Range : 0 O to 1 MO
- DC Voltage:
• Maximum Input : 50 V on any range
• Common Mode Rejection : 140 dB at 50 Hz or 60 Hz (1 kO unbalance in LOW lead) ±50 V peak maximum
- Resistance
• Max. Lead Resistance (4-wire ohms) : 10 O per lead for 100 O and 1 kO ranges.
- Digital I/O : Absolute voltage range : –4 V to 30 V

General Specifications
- Mains voltage 100 V setting:
• 90 V to 110 V
• 120 V setting 108 V to 132 V
• 220 V setting 198 V to 242 V
• 240 V setting 216 V to 264 V
- Frequency : 47 Hz to 440 Hz
- Power consumption : 36 VA peak (24 W average)
- Environment temperature : Operating : 0°C to 50°C
- Full accuracy : 18°C to 28°C
- Storage : -20°C to 70°C
- Relative humidity (non-condensing) : Operating 0°C to 30°C <80%; 30°C to 50°C <50%
- Storage : -20°C to 70°C <95%
- Altitude : Operating : 2,000 m
- Storage : 12,000 m
- Vibration and shock : Complies with MIL-PRF-28800F Class 3